A new HVE 1 MV multi-element AMS system at IHEG-CAGS AMS lab in China

PSB-37
23 Oct 2024, 17:35
20m
Lobby and Hallway , 2nd Floor

Lobby and Hallway , 2nd Floor

Poster New and Upgraded Facilities Poster Session B

Speaker

Hui Zhang

Description

A 1 MV Accelerator Mass Spectrometer(AMS) dedicated to measuring 14C as well as capabilities to measure 10Be, 26Al, 41Ca, 129I, was installed at IHEG-CAGS(Institute of Hydrogeology and Environmental Geology, Chinese Academy of Geological Sciences)in 2023. This multi-element AMS was developed by HVE and the results of an acceptance test demonstrated that its ability to achieve precision for 14C,10Be,26Al and 129I to 0.13,1.25,2.61 and 0.39% levels,with background values of 5.4×10-16, 1.4×10-15, 8.5×10-16 and 6.4×10-14 respectively.

Student Submission No

Primary authors

Presentation materials

There are no materials yet.